PULLUP BTB Adjustable Fixation System for Ligament Reconstruction, PULLUP CLIP Adjustable Fixation System for Ligament Reconstruction, PULLUP TEX CLIP Adjustable Fixation System for Ligament Reconstruction, BT LOOP Adjustable Fixation System for Ligament Reconstruction
Fastener, Fixation, Nondegradable, Soft Tissue. 510(k) K202193, from S.B.M. Sas (Science & Bio Materials) (Lourdes, FR). Product code MBI, class 2.
As of Nov 18, 2026, K202193 is an FDA 510(k) held by S.B.M. Sas (Science & Bio Materials), substantially equivalent on Oct 16, 2020 after 72 days, with no recalls tied to it in FDA's recall database.
- Review time: 72 days
- Recalls: 0
- Cited as a predicate by: 1 device
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe PULLUP® BTB, BT LOOP, PULLUP® CLIP and PULLUP® TEX CLIP devices are designed to be used as cortical fixation for ACL reconstruction.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- PULLUP® Adjustable Juxtacortical Fixation DeviceK151004 · Sbm Sciences For Bio Materials · Nov 17, 2015
- Force Fiber Blue Ultra High Molecular Weight Polyethylene Non Absorbable SuturesK092533 · Beere Precision Medical Instruments, Kmedic, Telef · Sep 15, 2009
- Sutures Polyprop., Polyetheline, Nylon & SilkK930739 · Deknatel, Inc. · Jul 26, 1994
Cited as a predicate
1 later 510(k) summary names this device as a predicate- PULLUP®; PULLUP® CLIP; BT LOOP®; PULLUP® TEX CLIP; RIGIDLOOP™ T; RIGIDLOOP™ Clip; RIGIDLOOP™ BTB; RIGIDLOOP™ Suture LoopK250844 · Science & Bio Materials (S.B.M.) Sas · Jun 2, 2025
Other MBI clearances
The newest 510(k)s for the same product code; their median review was 30 days; see every MBI device- Arthrex FiberButtonK263477 · Arthrex, Inc. · Oct 27, 2026
- Q-FIX KNOTLESS All-Suture Anchor, 1.8mm, with one ULTRABRAID Suture, Blue; Q-FIX KNOTLESS All-Suture Anchor, 1.8mm,…K263348 · Smith & Nephew, Inc. · Oct 23, 2026
- TeKBrace Knotless AnchorK263352 · TheraMicro · Oct 21, 2026
- Integrity Retractable Implant Inserter (6000127)K261869 · Anika Therapeutics, Inc. · Sep 17, 2026
- Aevumed SUPERCYNCH™ Button SystemK253065 · Aevumed, Inc. · Sep 2, 2026
- ArthroTAK Tendon Anchor KitK262809 · ArthroTAK, LLC · Sep 2, 2026
- Iconix Knotless TK262677 · Riverpoint Medical · Aug 27, 2026
- Grappler Suture Anchor SystemK262047 · Paragon 28, Inc. · Aug 3, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Aug 5, 2020
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3040
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files