Arca
Skip to content
Tools
Tools

Servo-u Ventilator System 4.1, Servo-n Ventilator System 4.1, Servo-u MR Ventilator System 4.1

Ventilator, Continuous, Facility Use. 510(k) K201874, from Maquet Critical Care AB (Solna, SE). Product code CBK, class 2.

As of Nov 18, 2026, K201874 is an FDA 510(k) held by Maquet Critical Care AB, substantially equivalent on Apr 20, 2021 after 287 days, with 5 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Apr 20, 2021
  • Review time: 287 days
  • Recalls: 5

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Servo-u Ventilator System is: • intended for respiratory support, monitoring and treatment of neonatal, pediatric and adult patients • to be used only by healthcare providers • to be used only in professional healthcare facilities and for transport within these facilities The Servo-n Ventilator System is: • intended for respiratory support, monitoring and treatment of neonatal and pediatric…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K201874, newest first; each opens FDA's recall record
Jun 30, 2026Z-2963-2026The module did not undergo electrical safety tests and functional tests during production.Open, Classified
May 30, 2025Z-2014-2025Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.Open, Classified
May 30, 2025Z-2013-2025Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.Open, Classified
May 30, 2025Z-2015-2025Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.Open, Classified
Jul 20, 2022Z-1560-2022Ventilators may generate a combination of alarms which may result in loss of communication, technical alarms being triggered, and loss of ventilation. If the device stops ventilating due to this issue, it may lead to…Open, Classified

Other CBK Clearances

The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device

Record

Type
Traditional
Received
Jul 7, 2020
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files