Tempus Pro Patient Monitor
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K201746, from Remote Diagnostic Technologies Limited (Basingstoke, GB). Product code MHX, class 2.
As of Nov 18, 2026, K201746 is an FDA 510(k) held by Remote Diagnostic Technologies Limited, substantially equivalent on Sep 18, 2020 after 84 days, with 7 recalls tied to it in FDA's recall database.
- Substantially equivalent: Sep 18, 2020
- Review time: 84 days
- Recalls: 7
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel, for the attended or unattended monitoring of single or multiple vital signs in clinical and pre-hospital care applications.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Tempus Pro Patient MonitorK173768 · Remote Diagnostic Technologies , Ltd. · Jan 10, 2018
- Infinity CentralStation WideK151860 · Draeger Medical Systems, Inc. · Sep 2, 2015
Recalls
Recalls FDA ties to K201746, newest first; each opens FDA's recall record| Nov 26, 2025 | Z-1027-2026Patient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for accuracy. If healthcare providers base clinical judgment on unvalidated… | Open, Classified |
| May 30, 2025 | Z-2175-2025Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead to… | Open, Classified |
| Apr 16, 2025 | Z-1666-2025software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope. | Open, Classified |
| Apr 15, 2025 | Z-1729-2025Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May reboot… | Open, Classified |
| Aug 24, 2023 | Z-2625-2023Full screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from monitors, informing users a 60-100 second shutdown/restart are required.… | Open, Classified |
| Feb 23, 2023 | Z-1321-2023During internal testing, Regulatory Compliance issues were identified regarding the Monitor and Power Supplies for fluid ingress and basic safety issues. | Open, Classified |
| May 25, 2021 | Z-1997-2021The Tempus Pro (Trizeps 7 only) when used in combination with a specified laryngoscope device can cause two error types when subsequently unplugged. | Open, Classified |
Other MHX Clearances
The newest 510(k)s for the same product code; their median review was 226 days; see every MHX device- Radius VSM and AccessoriesK260676 · Masimo Corporation · Oct 14, 2026
- Carescape ONE; Carescape Dock F0; TruSignal ECG MM; TruSignal PRES MM; TruSignal TEMP MM; TruSignal SpO2 MM 1.8M;…K260656 · GE Healthcare Finland Oy · Sep 1, 2026
- Carescape Canvas 1000; Carescape Canvas Smart DisplayK260343 · GE Healthcare Finland Oy · Aug 28, 2026
- Infinity® Acute Care System (IACS) VG8.1K260579 · Draeger Medical Systems, Inc. · Aug 20, 2026
- Telemetry Monitor (iT30, iT50)K251508 · Edan Instruments, Inc. · Aug 6, 2026
- Monitor B105M; Monitor B125M; Monitor B155M; Monitor B105P; Monitor B125PK253277 · Ge Medical Systems Information Technologies, Inc. · Jul 24, 2026
- Radius VSM and AccessoriesK254209 · Masimo Corporation · Jun 4, 2026
- M12 Telemetry SystemK252840 · Global Instrumentation, LLC · May 29, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Jun 26, 2020
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1025
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files