Arca
Skip to content
Tools
Tools

YSIO X.pree

System, X-Ray, Stationary. 510(k) K201670, from Siemens Medical Solutions USA, Inc. (Malvern, PA). Product code KPR, class 2.

As of Nov 20, 2026, K201670 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Oct 21, 2020 after 124 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 124 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The device is a digital X-ray system to generate X-ray images from the whole body including the skull, chest, abdomen, and extremities. The acquired images support medical professionals to make diagnostic and/or therapeutic decisions. Generic clinical benefits of radiographic examinations within the intended use are applicable for this device. YSIO X.pree is not for mammography examinations.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • YSIO X.preeK233543 · Siemens Medical Solutions · May 21, 2024

Recalls

Recalls FDA ties to K201670, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 7, 2022Z-0730-2022For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray…Open
Mar 10, 2021Z-1346-2021A patient image was assigned to a different patient during image recovery process. In the unlikely event of interrupted connection to a MAX detector during initiated image recovery, instead of the currently acquired…Terminated

Other KPR clearances

The newest 510(k)s for the same product code; their median review was 195 days; see every KPR device

Record

Type
Traditional
Received
Jun 19, 2020
Review panel
Radiology
Regulation
21 CFR 892.1680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files