Device, Positive Pressure Breathing, Intermittent. 510(k) K200988, from Hill-Rom Services Pte, Ltd. (Singapore, SG). Product code NHJ, class 2.

As of Nov 20, 2026, K200988 is an FDA 510(k) held by Hill-Rom Services Pte, Ltd., substantially equivalent on May 26, 2020 after 41 days, with 10 recalls tied to it in FDA's recall database.

  • Review time: 41 days
  • Recalls: 10
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Maximus™ System provides features of both the Synclara™ System and the Volara™ System. The Maximus™ System, when used as a Synclara™ Cough System is intended for use on patients who are unable to cough or clear secretions effectively due to reduced peak cough expiratory flow or respiratory muscle weakness.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K200988, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 3, 2026Z-3257-2026Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.Open
May 21, 2026Z-2549-2026, Z-2551-2026, Z-2550-2026, Z-2548-2026There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization,…Open
Feb 18, 2026Z-1637-2026Urgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or…Open
May 30, 2024Z-2322-2024, Z-2321-2024Potential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter. The handset plug is required to ensure proper operation and ventilator gas flow.Open
Apr 26, 2022Z-1138-2022, Z-1139-2022There is a potential of reduced oxygenation or pneumothorax/barotrauma to occur when the Volara system is used with in-line ventilator in home care environment.Terminated

Other NHJ clearances

The newest 510(k)s for the same product code; their median review was 273 days; see every NHJ device

Record

Type
Special
Received
Apr 15, 2020
Review panel
Anesthesiology
Regulation
21 CFR 868.5905

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files