Neurological Stereotaxic Instrument. 510(k) K200511, from Medtech S.A (Montpellier, FR). Product code HAW, class 2.
As of Nov 18, 2026, K200511 is an FDA 510(k) held by Medtech S.A, substantially equivalent on May 29, 2020 after 88 days, with 1 recall tied to it in FDA's recall database.
- Review time: 88 days
- Recalls: 1
- Cited as a predicate by: 3 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic neurosurgery may be appropriate.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- ROSA ONE Brain ApplicationK182417 · Medtech S.A · Feb 7, 2019
Cited as a predicate
3 later 510(k) summaries name this device as a predicate- Geniant Cranial (Navigated Neurosurgical Positioning Robot)K241333 · Koh Young Technology, Inc. · Jan 17, 2025
- Sinobot X1K220072 · Sinovation (Beijing) Medical Technology Co., Ltd. · Jun 18, 2023
- ROSA ONE Brain applicationK214065 · Medtech S.A.S · May 4, 2022
Recalls
Recalls FDA ties to K200511, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Started | Recall | Status |
|---|---|---|
| Sep 22, 2021 | Z-0118-2022The firm has become aware of a software anomaly affecting ROSA One 3.1 Brain application which led to the inaccurate placement of an electrode during surgery. The firm has received 3 global complaints related to the… | Open |
Other HAW clearances
The newest 510(k)s for the same product code; their median review was 162 days; see every HAW device- Stereotactic Ultralight System SUSyK261157 · inomed Medizintechnik GmbH · Nov 6, 2026
- StealthStation™ S8 Cranial v2.3.0K262716 · Medtronic Navigation · Nov 6, 2026
- NeuroAlignK261997 · Medivis, Inc. · Oct 26, 2026
- Disposable Reflective Marker Spheres (AM-BK115S, AM-BL130S)K262214 · Guangzhou Aimooe Technology Co., Ltd. · Oct 15, 2026
- Geniant Cranial (Navigated Neurosurgical Positioning Robot)K253856 · Koh Young Technology, Inc. · Sep 10, 2026
- Leksell® Vantage Stereotactic SystemK261932 · Elekta Solutions AB · Aug 5, 2026
- Zeta Navigation SystemK253663 · Zeta Surgical, Inc. · May 1, 2026
- Stealth AXiS Cranial clinical applicationK253379 · Medtronic Navigation, Inc. · Mar 26, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Mar 2, 2020
- Review panel
- Neurology
- Regulation
- 21 CFR 882.4560
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files