Arca
Skip to content
Tools
Tools

Catalyst+, Catalyst+ HD

Accelerator, Linear, Medical. 510(k) K200435, from C-Rad Positioning AB (Uppsala, SE). Product code IYE, class 2.

As of Nov 20, 2026, K200435 is an FDA 510(k) held by C-Rad Positioning AB, substantially equivalent on May 7, 2020 after 73 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 73 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Catalyst+ system and accessories are intended for use in radiation therapy clinics with treatment equipment, to provide: • Interactive guidance for setup and positioning of patients for radiation therapy • Intra-fraction motion detection during radiation therapy • Respiratory gating during radiation therapy The device is not intended for use in MR-linac systems.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • CatalystK113276 · C-Rad Positioning AB · Feb 27, 2012

Recalls

Recalls FDA ties to K200435, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 9, 2025Z-1092-2026Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the…Open
Apr 22, 2025Z-1858-2025Software issue with scanning equipment that can results in the filed rotation not being applied correctly.Open

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Feb 24, 2020
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.