Biograph mMR with syngo MR E11P-AP01 system software
Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance. 510(k) K200213, from Siemens Medical Solutions USA, Inc. (Malvern, PA). Product code OUO, class 2.
As of Nov 20, 2026, K200213 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on May 11, 2020 after 104 days, with 1 recall tied to it in FDA's recall database.
- Review time: 104 days
- Recalls: 1
- Cited as a predicate by: 1 device
The Siemens MR-PET system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information, acquired simultaneously and isocentrically.
- Biograph mMR with syngo MR E11P system softwareK163234 · Siemens Medical Solutions USA, Inc. · Feb 28, 2017
- 1.5T 32-Channel Head Coil and 3T 32-Channel Head CoilK072909 · Siemens AG Medical Solutions · Oct 31, 2007
- MAGNETOM Avanto, MAGNETOM Verio with software syngo MR E11DK181613 · Siemens Medical Solutions USA, Inc. · Nov 6, 2018
- BIOGRAPH OneK253023 · Siemens Healthineers AG · Jan 15, 2026
| Started | Recall | Status |
|---|---|---|
| Aug 28, 2025 | Z-2654-2025There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a… | Open |
- uPMR ZenithK261676 · Shanghai United Imaging Healthcare Co., Ltd. · Oct 28, 2026
- Cubresa BrainPETK253963 · Cubresa, Inc. · Jul 17, 2026
- BIOGRAPH OneK253023 · Siemens Healthineers AG · Jan 15, 2026
- uPMR 790K234154 · Shanghai United Imaging Healthcare Co., Ltd. · May 24, 2024
- uPMR 790K222540 · Shanghai United Imaging Healthcare Co., Ltd. · Nov 14, 2022
- Signa Pet/MrK213709 · Ge Medical Systems, LLC · Jan 20, 2022
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 28, 2020
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1200
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files