Arca
Skip to content

PERLA® TL posterior osteosynthesis system

Thoracolumbosacral Pedicle Screw System. 510(k) K193396, from Spineart (Plan Les Ouates, CH). Product code NKB, class 2.

As of Nov 17, 2026, K193396 is an FDA 510(k) held by Spineart, substantially equivalent on Feb 5, 2020 after 61 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Feb 5, 2020
  • Review time: 61 days
  • Recalls: 2
  • Cited as a predicate by: 6 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The PERLA® TL posterior osteosynthesis system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K193396, newest first; each opens FDA's recall record
Mar 15, 2024Z-1758-2024Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but…Open, Classified
Apr 9, 2021Z-1975-2021Due to external labelling error/mix-up. Affected product external labels list the incorrect rod size.Terminated

Other NKB Clearances

The newest 510(k)s for the same product code; their median review was 90 days; see every NKB device

Record

Type
Traditional
Received
Dec 6, 2019
Review panel
Orthopedic
Regulation
21 CFR 888.3070

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.