Smart Correction System
Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component. 510(k) K193368, from Wishbone Medical, Inc. (Warsaw, IN). Product code KTT, class 2.
As of Nov 19, 2026, K193368 is an FDA 510(k) held by Wishbone Medical, Inc., substantially equivalent on Sep 18, 2020 after 288 days, with 1 recall tied to it in FDA's recall database.
- Review time: 288 days
- Recalls: 1
- Cited as a predicate by: 3 devices
The Smart Correction® System is indicated for pediatric subpopulations (excluding newborns) and adults for the following: • Joint contracture resulting in loss of range of motion • Fractures and disease which generally may result in joint contractures or loss of range of motion • Fractures requiring distraction • Open and closed fracture fixation, including fractures of long bones (intracapsular…
- Taylor Spatial Frame External Fixation SystemK970748 · Smith & Nephew, Inc., Orthopaedic Div. · May 9, 1997
- Smith & Nephew Circular Fixation SystemK093047 · Smith & Nephew, Inc. · Sep 27, 2010
- Orthofix Tl-Hex True Lok Hexalobe System (Tl-Hex)K141078 · Orthofix Srl · Sep 2, 2014
- Orthofix True Lok Hexapod System VI.3K143125 · Orthofix Srl · Dec 10, 2014
- Orthofix TL-HEX True Lok Hexapod System (TL-HEX) V1.4K152171 · Orthofix Srl · Sep 28, 2015
- Orthofix TrueLok Hexapod System (TL-HEX) V2.0K170650 · Orthofix Srl · May 10, 2017
- The Response Ortho Smart Fixator - Hexapod SystemK252625 · Response Ortho Solutions, LLC · Nov 20, 2025
- Smart Correction System (HA Half Pins)K231461 · Wishbone Medical, Inc. · Jun 16, 2023
- Smart Correction System Rings and Compatible HA-Coated Half PinsK221366 · Wishbone Medical, Inc. · Jun 6, 2022
| Started | Recall | Status |
|---|---|---|
| Jan 4, 2023 | Z-1102-2023Firm identified that user may be unable to adjust the length of the Smart Correction Dual Joint Express Struts due to the collar release pin backing out. This may lead to a delay in correction. | Open |
- Response Ortho Smart Fixator - Hexapod SystemK263083 · Response Ortho Solutions, LLC · Oct 2, 2026
- MAXFRAME AUTOSTRUT™ with Bluetooth Multi-Axial Correction SystemK254087 · Synthes (USA) Products, LLC · Jun 11, 2026
- DYNEX® External Fixation SystemsK260900 · Vilex, LLC · Apr 17, 2026
- QuikFix External Fixator: Knee-Spanning PackK260112 · Stabiliz Orthopaedics, Inc. · Mar 23, 2026
- Excelsior SystemK253291 · Blue Ocean Global · Jan 20, 2026
- Hoffmann LRF SystemK253202 · Stryker GmbH · Dec 19, 2025
- Monkey Rings External Ring Fixation SystemK253613 · Paragon 28, Inc. · Dec 19, 2025
- Medusa Orthopedics Boa External Fixation SystemK252555 · Medusa Orthopedics, LLC · Nov 25, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 5, 2019
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3030
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files