Arca
Skip to content
Tools
Tools

Puritan Bennett 980 Series Ventilator System

Ventilator, Continuous, Facility Use. 510(k) K193056, from Covidien, LLC (Boulder, CO). Product code CBK, class 2.

As of Nov 20, 2026, K193056 is an FDA 510(k) held by Covidien, LLC, substantially equivalent on Nov 20, 2020 after 385 days, with 17 recalls tied to it in FDA's recall database.

  • Review time: 385 days
  • Recalls: 17
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Puritan Bennett 980 Series Ventilator System is designed for use on patient population sizes from Neonatal (NICU) through Adult who require respiratory support or mechanical ventilation and weigh a minimum of 0.3kg (0.6lb).

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K193056, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 25, 2022Z-0966-2022The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially…Open
Nov 4, 2021Z-0347-2022, Z-0340-2022, Z-0346-2022, Z-0352-2022, Z-0343-2022, Z-0341-2022, Z-0345-2022, Z-0339-2022, Z-0353-2022, Z-0350-2022, Z-0342-2022, Z-0348-2022, Z-0349-2022, Z-0344-2022, Z-0338-2022, Z-0351-2022Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.Terminated

Other CBK clearances

The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device

Record

Type
Traditional
Received
Nov 1, 2019
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files