Elekta Unity

Accelerator, Linear, Medical. 510(k) K192482, from Elekta Limited (Crawley, GB). Product code IYE, class 2.

As of Nov 20, 2026, K192482 is an FDA 510(k) held by Elekta Limited, substantially equivalent on Dec 5, 2019 after 86 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 86 days
  • Recalls: 3
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with the compatible Treatment Planning and Oncology Information Systems.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K192482, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 21, 2020Z-2297-2020In TSM (Treatment Session Manager) Motion Monitoring workflows, under certain conditions there can be a mismatch between the contour data overlay with respect to the motion monitoring images of the monitored structure.Terminated
Apr 30, 2020Z-2289-2020potential risk that Legionella may be present in the Unity machine room, specifically the heat exchanger condensation collection tray.Terminated
Dec 18, 2019Z-2473-2020There have been reports of excessive heating in the MRL Gradient Connection Module.Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Sep 10, 2019
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files