PhoeniX
System, X-Ray, Mobile. 510(k) K192011, from Sedecal., Sa. (Algete, ES). Product code IZL, class 2.
As of Nov 18, 2026, K192011 is an FDA 510(k) held by Sedecal., Sa., substantially equivalent on Sep 25, 2019 after 58 days, with no recalls tied to it in FDA's recall database.
- Substantially equivalent: Sep 25, 2019
- Review time: 58 days
- Recalls: 0
- Cited as a predicate by: 4 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFIntended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- RadPRO® Mobile 40kW; RadPRO® Mobile 40kW FLEXPLUS, Model SM-40HF-B-D-VIRK161345 · Sedecal., Sa. · Jul 6, 2016
- Digital Radiography CXDI-401C WirelessK133693 · Canon, Inc. · Jul 1, 2014
- Digital Radiography CXDI-701C Wireless, Digital Radiography CXDI-70IG Wireless, Digital Radiography CXDI-801C Wireless,K131106 · Canon, Inc. · Jul 3, 2013
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- MOVIX DReamyK212515 · Stephanix · Nov 3, 2021
- PhoenixK212291 · Sedecal., Sa. · Sep 14, 2021
- Diagnostic X-ray System/PLX5200AK212134 · Nanjing Perlove Medical Equipment Co., Ltd. · Aug 25, 2021
- Soltus 500K192936 · Sedecal., Sa. · Nov 15, 2019
Other IZL Clearances
The newest 510(k)s for the same product code; their median review was 166 days; see every IZL device- REMEX-T(K)100CK261943 · Remedi, Inc. · Oct 2, 2026
- XliteK260236 · Gemss Healthcare Co., Ltd. · Aug 3, 2026
- Portable DR Imaging SystemK253058 · Shantou Institute of Ultrasonic Instruments Co., Ltd. · Jun 18, 2026
- RAYMO Mobile X-ray System (Model: RAYMO)K260085 · DRGEM Corporation · May 14, 2026
- Lux HD Detectors, DRX-LC Detector, DRX-Revolution Mobile X-Ray SystemK253185 · Carestream Health · Mar 11, 2026
- AirRay-mini 80A, AirRay-mini 90A, AirRay-mini 80F, AirRay-mini 90F (AirRay-mini)K253244 · Poskom Co., Ltd. · Feb 23, 2026
- Airbile-100K250976 · Poskom Co., Ltd. · Dec 19, 2025
- Mobile X-ray unit (!M1)K251710 · Solutions For Tomorrow AB · Oct 24, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Abbreviated
- Received
- Jul 29, 2019
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1720
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files