Accessories, Pump, Infusion. 510(k) K191910, from B. Braun Melsugen AG (Melsungen, DE). Product code MRZ, class 2.
As of Nov 20, 2026, K191910 is an FDA 510(k) held by B. Braun Melsugen AG, substantially equivalent on Mar 12, 2020 after 239 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 239 days
- Recalls: 8
The SpaceStation MRI is a MRI (Magnetic Resonance Imaging) System Rack for the operation of Space Infusion Pumps during MRI examinations (MRI procedures) of adult, pediatric or neonatal patients. The product is intended to be used by qualified healthcare professionals.
- Perfusor Space Infusion Syringe Pump SystemK062699 · B.Braun Medical, Inc. · Oct 26, 2006
- Infusomat Space Volumetric Infusion Pump SystemK062700 · B.Braun Medical, Inc. · Oct 26, 2006
- Mri-CaddyK030323 · Mammendorfer Institute For Physics & Medicine GmbH · Apr 2, 2003
| Aug 7, 2024 | Z-2972-2024Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream… | Open |
| Jun 14, 2024 | Z-2434-2024, Z-2437-2024, Z-2433-2024, Z-2435-2024Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with… | Open |
| Sep 21, 2023 | Z-0123-2024, Z-0124-2024Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to… | Open |
| Feb 7, 2022 | Z-0819-2022Units not properly inspected prior to release to the market, may not function as indicated in the product labeling, resulting in syringe detection and infusion accuracy that can result in delays in therapy,… | Terminated |
- UTC 3mL Medication CartridgeK230718 · Gilero, LLC · Jun 15, 2023
- Elite Biomedical Solutions Dose Request CordK220695 · Elite Biomedical Solutions · Apr 29, 2022
- RG 3ml Medication CartridgeK201634 · Chengdu Shifeng Medical Technology Co., Ltd. · Mar 26, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 17, 2019
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5725
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files