Attune Revision LPS Inserts
Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer. 510(k) K191779, from Depuy(Ireland) (Cork, IE). Product code KRO, class 2.
As of Nov 17, 2026, K191779 is an FDA 510(k) held by Depuy(Ireland), substantially equivalent on Oct 8, 2019 after 98 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Oct 8, 2019
- Review time: 98 days
- Recalls: 4
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe ATTUNE Revision LPS Insert is intended for use in prosthetic constructs for replacement of distal and/or total femur and resurfacing of the proximal tibia, especially in cases that require extensive resection and replacement.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Lps Universal Tibial Hinge Inserts, Model 198727012-198727331K091453 · DePuy Orthopaedics, Inc. · Aug 12, 2009
- Depuy Attune Knee SystemK101433 · DePuy Orthopaedics, Inc. · Dec 10, 2010
- Depuy Sttune (Tm) Ps Knee SystemK111433 · DePuy Orthopaedics, Inc. · Aug 30, 2011
- ATTUNE Revision Knee SystemK160700 · Depuy(Ireland) · Jun 10, 2016
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- ATTUNE Medial Stabilized (MS) Fixed Bearing (FB) InsertK211609 · Depuy Ireland UC · Aug 25, 2021
Recalls
Recalls FDA ties to K191779, newest first; each opens FDA's recall record| Feb 15, 2023 | Z-1265-2023The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to… | Open, Classified |
| Feb 15, 2023 | Z-1263-2023The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to… | Open, Classified |
| Feb 15, 2023 | Z-1266-2023The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to… | Open, Classified |
| Feb 15, 2023 | Z-1264-2023The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to… | Open, Classified |
Other KRO Clearances
The newest 510(k)s for the same product code; their median review was 87 days; see every KRO device- Triathlon® Knee System; Triathlon® Hinge Knee System; Scorpio® Total Knee SystemK261898 · Howmedica Osteonics Corp. dba Stryker Orthopaedics · Aug 6, 2026
- Triathlon® Tritanium® Asymmetric Patella; Triathlon® Tritanium® Symmetric Patella;…K252898 · Howmedica Osteonics Corp., Dba Stryker Orthopaedics · Dec 15, 2025
- Stem Extension Line (USTAR II System)K252303 · United Orthopedic Corporation · Aug 22, 2025
- Triathlon® Hinge Knee SystemK251665 · Howmedica Osteonics Corp., Dba Stryker Orthopaedics · Jul 25, 2025
- ATTUNE Revision Hinge KneeK242871 · Depuy Ireland UC · Nov 21, 2024
- EVOLUTION® Hinge Knee System;GUARDIAN limb Salvage System;DCW Modular Distal Femoral…K240043 · Microport Orthopedics, Inc. · Apr 1, 2024
- ATTUNE Total Knee System, ATTUNE Cementless Knee System, LPS Limb Preservation System,…K233980 · Depuy Ireland UC · Mar 21, 2024
- EVOLUTION® Hinge Knee SystemK230563 · Microport Orthopedics, Inc. · Sep 27, 2023
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 2, 2019
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3510
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files