Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer. 510(k) K191779, from Depuy(Ireland) (Cork, IE). Product code KRO, class 2.

As of Nov 17, 2026, K191779 is an FDA 510(k) held by Depuy(Ireland), substantially equivalent on Oct 8, 2019 after 98 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Oct 8, 2019
  • Review time: 98 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The ATTUNE Revision LPS Insert is intended for use in prosthetic constructs for replacement of distal and/or total femur and resurfacing of the proximal tibia, especially in cases that require extensive resection and replacement.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K191779, newest first; each opens FDA's recall record
Feb 15, 2023Z-1265-2023The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to…Open, Classified
Feb 15, 2023Z-1263-2023The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to…Open, Classified
Feb 15, 2023Z-1266-2023The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to…Open, Classified
Feb 15, 2023Z-1264-2023The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to…Open, Classified

Other KRO Clearances

The newest 510(k)s for the same product code; their median review was 87 days; see every KRO device

Record

Type
Traditional
Received
Jul 2, 2019
Review panel
Orthopedic
Regulation
21 CFR 888.3510

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files