Arca
Skip to content
Tools
Tools

Tablo Hemodialysis System, Tablo Cartridge

Dialyzer, High Permeability With Or Without Sealed Dialysate System. 510(k) K190793, from Outset Medical, Inc. (San Jose, CA). Product code KDI, class 2.

As of Nov 19, 2026, K190793 is an FDA 510(k) held by Outset Medical, Inc., substantially equivalent on Dec 12, 2019 after 259 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 259 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Tablo® Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician’s prescription, with a trained individual available as needed who is considered competent in the use of the device by the prescribing physician.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K190793, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 24, 2023Z-1351-2023An observed trend of high conductivity dialysate alarms in a hemodialysis system as a result of introducing a software tool into the manufacturing process.Open
Jan 20, 2022Z-0813-2022Due to a component in the hemodialysis console there is the possibility of heat-related damage that may occur with the device.Open

Other KDI clearances

The newest 510(k)s for the same product code; their median review was 189 days; see every KDI device

Record

Type
Traditional
Received
Mar 28, 2019
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files