SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM Definition AS Open, SOMATOM Edge Plus, SOMATOM Definition AS/AS+, SOMATOM Confidence
System, X-Ray, Tomography, Computed. 510(k) K190578, from Siemens Medical Solutions USA, Inc. (Malvern, PA). Product code JAK, class 2.
As of Nov 17, 2026, K190578 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Jun 27, 2019 after 113 days, with 13 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jun 27, 2019
- Review time: 113 days
- Recalls: 13
- Cited as a predicate by: 5 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThis computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM…K173630 · Siemens Medical Solutions USA, Inc. · Mar 30, 2018
- SOMATOM Edge PlusK173607 · Siemens Medical Solutions USA, Inc. · Mar 21, 2018
Cited as a Predicate
5 later 510(k) summaries name this device as a predicate- SOMATOM Edge Plus, SOMATOM Confidence, SOMATOM Definition Edge, SOMATOM Definition…K230421 · Siemens Medical Solutions USA, Inc. · Jun 16, 2023
- NAEOTOM Alpha, Scan&GOK211591 · Siemens Medical Solutions USA, Inc. · Sep 30, 2021
- SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All, go.Top, SOMATOM X. Platform…K211373 · Siemens Medical Solutions USA, Inc. · Aug 27, 2021
- SOMATOM Go Platform – SOMATOM go.Now, go.Up, go.All, go.Top, go.Sim, go.Open Pro, Scan &…K192061 · Siemens Medical Solutions USA, Inc. · Nov 21, 2019
- SOMATOM X.citeK191891 · Siemens Medical Solutions USA, Inc. · Nov 6, 2019
Recalls
Recalls FDA ties to K190578, newest first; each opens FDA's recall record| Nov 19, 2021 | Z-0474-2022Software version VB20_SP5 my lead to a relevant degradation of head image quality, a potential risk to patients is misdiagnosis. | Open, Classified |
| Nov 19, 2021 | Z-0472-2022Software version VB20_SP5 my lead to a relevant degradation of head image quality, a potential risk to patients is misdiagnosis. | Open, Classified |
| Nov 19, 2021 | Z-0473-2022Software version VB20_SP5 my lead to a relevant degradation of head image quality, a potential risk to patients is misdiagnosis. | Open, Classified |
| Nov 19, 2021 | Z-0475-2022Software version VB20_SP5 my lead to a relevant degradation of head image quality, a potential risk to patients is misdiagnosis. | Open, Classified |
| Aug 30, 2021 | Z-0286-2022software syngo.CT VB20 in the installed base, with or without syngo.CT software versions VB20_SP1, SP2, SP3, or SP4, may result in workflow interruptions resulting in the delay in diagnosis, the necessity for patient… | Terminated |
| Aug 30, 2021 | Z-0283-2022software syngo.CT VB20 in the installed base, with or without syngo.CT software versions VB20_SP1, SP2, SP3, or SP4, may result in workflow interruptions resulting in the delay in diagnosis, the necessity for patient… | Terminated |
| Aug 30, 2021 | Z-0288-2022software syngo.CT VB20 in the installed base, with or without syngo.CT software versions VB20_SP1, SP2, SP3, or SP4, may result in workflow interruptions resulting in the delay in diagnosis, the necessity for patient… | Terminated |
| Aug 30, 2021 | Z-0287-2022software syngo.CT VB20 in the installed base, with or without syngo.CT software versions VB20_SP1, SP2, SP3, or SP4, may result in workflow interruptions resulting in the delay in diagnosis, the necessity for patient… | Terminated |
| Aug 30, 2021 | Z-0284-2022software syngo.CT VB20 in the installed base, with or without syngo.CT software versions VB20_SP1, SP2, SP3, or SP4, may result in workflow interruptions resulting in the delay in diagnosis, the necessity for patient… | Terminated |
| Aug 30, 2021 | Z-0285-2022software syngo.CT VB20 in the installed base, with or without syngo.CT software versions VB20_SP1, SP2, SP3, or SP4, may result in workflow interruptions resulting in the delay in diagnosis, the necessity for patient… | Terminated |
| Aug 30, 2021 | Z-0289-2022software syngo.CT VB20 in the installed base, with or without syngo.CT software versions VB20_SP1, SP2, SP3, or SP4, may result in workflow interruptions resulting in the delay in diagnosis, the necessity for patient… | Terminated |
| Jul 25, 2019 | Z-2440-2019SOMATOM Definition Edge, SOMATOM Definition AS, SOMATOM Definition Flash with software syngo.CT with Software version VA48A_SP5 may result in scanning workflow interruptions and unexpected user notifications. and may… | Terminated |
| Jul 3, 2019 | Z-2270-2019Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort. | Terminated |
Other JAK Clearances
The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device- AVIEW Aorta MetricsK260391 · Coreline Soft Co., Ltd. · Oct 16, 2026
- CAS-One IRK261225 · Cascination AG · Sep 29, 2026
- BodyTom 64 (NL4100); OmniTom Elite (NL5000)K261125 · Samsung HME America, Inc., dba NeuroLogica Corp. · Sep 22, 2026
- SCENARIA View Phase 6.0K261229 · Fujifilm Corporation · Sep 18, 2026
- FCT iStream Phase 2K253836 · Fujifilm Corporation · Sep 4, 2026
- BodyTom Elite (NL4000)K253450 · NeuroLogica Corp., a Samsung Company · Jul 28, 2026
- CardIQ SuiteK253770 · GE Medical Systems SCS · Jul 21, 2026
- Enhanced Boundary for PCCTK253560 · Ge Medical Systems, LLC · Jun 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Mar 6, 2019
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files