Arca
Skip to content
Tools
Tools

CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO

Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase. 510(k) K183394, from GE Healthcare Finland Oy (Helsinki, FI). Product code CCK, class 2.

As of Nov 20, 2026, K183394 is an FDA 510(k) held by GE Healthcare Finland Oy, substantially equivalent on May 1, 2019 after 145 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 145 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The CARESCAPE Respiratory Modules (E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX) are indicated for use with a host device for monitoring respiratory parameters (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate) and ventilatory parameters (airway pressure, flow and volume) of adult, pediatric and neonatal patients and gas exchange parameters (VCO2, VO2)…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K183394, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Sep 27, 2019Z-0756-2020, Z-0755-2020, Z-0754-2020, Z-0757-2020Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.Terminated

Other CCK clearances

The newest 510(k)s for the same product code; their median review was 224 days; see every CCK device

Record

Type
Special
Received
Dec 7, 2018
Review panel
Anesthesiology
Regulation
21 CFR 868.1400

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files