Arca
Skip to content
Tools
Tools

Bellavista

Ventilator, Continuous, Facility Use. 510(k) K183364, from Imtmedical AG (Buchs Sg, CH). Product code CBK, class 2.

As of Nov 18, 2026, K183364 is an FDA 510(k) held by Imtmedical AG, substantially equivalent on Sep 13, 2019 after 283 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 283 days
  • Recalls: 3

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The bellavista 1000/1000e ventilator is intended to provide positive pressure ventilatory support to adult and pediatric patients and optionally infant and neonatal patients by qualified, trained personnel under the direction of a physician.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K183364, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 5, 2023Z-1966-2023, Z-1967-2023Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops…Completed
Dec 20, 2021Z-0622-2022Potential cessation in ventilation can occur under specific conditions.Terminated

Other CBK clearances

The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device

Record

Type
Traditional
Received
Dec 4, 2018
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.