Coag-Sense Prothrombin Time (PT) / INR Monitoring Device
Test, Time, Prothrombin. 510(k) K183255, from Coagusense, Inc. (Fremont, CA). Product code GJS, class 2.
As of Nov 18, 2026, K183255 is an FDA 510(k) held by Coagusense, Inc., substantially equivalent on Feb 21, 2019 after 92 days, with 1 recall tied to it in FDA's recall database.
- Review time: 92 days
- Recalls: 1
- Cited as a predicate by: 1 device
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe Coag-Sense® Prothrombin Time (PT) / INR Monitoring System (Self-Test) is an in vitro diagnostic device that provides quantitative prothrombin time (PT) results, expressed in seconds and INR units. It uses fresh capillary whole blood.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Immedia Prothrombin Time SystemK050243 · Farallon Medial, Inc. · Jun 6, 2005
- Coagusense Self-Test Prothrombin Time/Inr Monitoring SystemK093243 · Coagusense, Inc. · Apr 7, 2010
Cited as a predicate
1 later 510(k) summary names this device as a predicate- Coag-Sense Prothrombin (PT) / INR Monitoring System for Patient Self-TestingK212779 · Coagusense, Inc. · Oct 5, 2022
Recalls
Recalls FDA ties to K183255, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Started | Recall | Status |
|---|---|---|
| May 25, 2022 | Z-1583-2022Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. | Open |
Other GJS clearances
The newest 510(k)s for the same product code; their median review was 299 days; see every GJS device- CoaguChek XS Plus SystemK253188 · Roche Diagnostics · Oct 24, 2025
- microINR SystemK251564 · Iline Microsystems, S.L. · Jul 21, 2025
- microINR SystemK243543 · Iline Microsystems, S.L. · Feb 27, 2025
- i-STAT PTplus Cartridge with the i-STAT 1 SystemK220282 · Abbott Laboratories · Jul 14, 2023
- microINR SystemK231711 · Iline Microsystems, S.L. · Jul 11, 2023
- STA- NeoPTimalK211485 · Diagnostica Stago S.A.S · Dec 23, 2022
- Coag-Sense Prothrombin (PT) / INR Monitoring System for Patient Self-TestingK212779 · Coagusense, Inc. · Oct 5, 2022
- HemosIL ReadiPlasTinK213426 · Instrumentation Laboratory CO · Aug 16, 2022
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 21, 2018
- Review panel
- Hematology
- Regulation
- 21 CFR 864.7750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files