Arca
Skip to content
Tools
Tools

Intraoperative, Single-Use, Sterile, Silicone Breast Sizer Motiva Implant Matrix

Mammary Sizer. 510(k) K183163, from Motiva USA, LLC (New York, NY). Product code MRD, class U.

As of Nov 24, 2026, K183163 is an FDA 510(k) held by Motiva USA, LLC, substantially equivalent on Feb 5, 2019 after 82 days, with no recalls tied to it in FDA's recall database.

  • Review time: 82 days
  • Recalls: 0
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Intraoperative, Single-use, Sterile, Silicone Breast Sizers Motiva Implant Matrix are temporary intraoperative placement devices used during breast augmentation or reconstruction procedures to assist the surgeon in determining the appropriate size of the long-term breast implant to use.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Other MRD clearances

The newest 510(k)s for the same product code; their median review was 98 days; see every MRD device

Record

Type
Traditional
Received
Nov 15, 2018

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.