Integra Salto Total Ankle System

Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer. 510(k) K182878, from Ascension Orthopedics (Austin, TX). Product code HSN, class 2.

As of Nov 18, 2026, K182878 is an FDA 510(k) held by Ascension Orthopedics, substantially equivalent on Dec 18, 2018 after 64 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 64 days
  • Recalls: 8
  • Cited as a predicate by: 1 device

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K182878, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Mar 8, 2019Z-1168-2019, Z-1173-2019, Z-1172-2019, Z-1175-2019, Z-1170-2019, Z-1169-2019, Z-1174-2019, Z-1171-2019The Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFUTerminated

Other HSN clearances

The newest 510(k)s for the same product code; their median review was 163 days; see every HSN device

Record

Type
Special
Received
Oct 15, 2018
Review panel
Orthopedic
Regulation
21 CFR 888.3110

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.