Sterilizable Internal Defibrillation Paddles for use with LIFEPAK defibrillators/monitors
Dc-Defibrillator, Low-Energy, (Including Paddles). 510(k) K182503, from Physio-Control, Inc. (Redmond, WA). Product code LDD, class 2.
As of Nov 21, 2026, K182503 is an FDA 510(k) held by Physio-Control, Inc., substantially equivalent on Jun 3, 2019 after 264 days, with 1 recall tied to it in FDA's recall database.
- Review time: 264 days
- Recalls: 1
- Cited as a predicate by: 1 device
The Sterilizable Internal Defibrillation Paddles are intended for use with LIFEPAK defibrillators to internally detect ECG rhythm and provide defibrillation or synchronized cardioversion directly to the surgically exposed heart within a sterile use environment.
- Low Energy DC - Defibrillator (Including Paddles)K895379 · Physio-Control Corp. · Feb 26, 1990
- Switched Internal PaddlesK203231 · Philips North America, LLC · Jan 11, 2021
| Started | Recall | Status |
|---|---|---|
| Aug 9, 2024 | Z-0142-2025Due to customer complaints and out of box failures, the firm is conducting a voluntary recall as a result of product failures (cracks and over-molding separation of the sterilizable internal defibrillation paddles. | Open |
- Switched Internal PaddlesK203231 · Philips North America, LLC · Jan 11, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 12, 2018
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.5300
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files