Stratum™ Foot Plating System
Plate, Fixation, Bone. 510(k) K182201, from Nextremity Solutions, Inc. (Warsaw, IN). Product code HRS, class 2.
As of Nov 18, 2026, K182201 is an FDA 510(k) held by Nextremity Solutions, Inc., substantially equivalent on Nov 16, 2018 after 94 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 94 days
- Recalls: 5
- Cited as a predicate by: 4 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe Nextremity Solutions Stratum™ Foot Plating System is a plate and screws construct indicated for fixation of fractures, osteotomies, non-unions, malunions and fusions of small bones and small bone segments, particularly in osteopenic bone.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Synthes (USA) Modular Mini Fragment Lcp SystemK063049 · Synthes (Usa) · Dec 18, 2006
- Msp Metatarsal Shortening SystemK140724 · Nextremity Solutions, Inc. · May 14, 2014
- Lapidus Plate, 3.5MM Inline Fusion Plate, 3.5MM Mtp Plate (Left and Right)K131670 · Biomet, Inc. · Jul 3, 2013
Cited as a predicate
4 later 510(k) summaries name this device as a predicate- InCore Subtalar SystemK213301 · Nextremity Solutions, Inc. · Dec 21, 2021
- CalcShift™ Displacement Calcaneal Osteotomy SystemK212979 · Nextremity Solutions, Inc. · May 20, 2022
- Stratum® Ankle Fusion Plating SystemK212640 · Nextremity Solutions, Inc. · Nov 18, 2021
- Stratum™ Reduced Size Foot Plating SystemK200785 · Nextremity Solutions, Inc. · Jul 27, 2020
Recalls
Recalls FDA ties to K182201, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Dec 16, 2019 | Z-1127-2020The product contains a different plate than indicated by the package labeling. The product is labeled as a right plate but contains a left plate. | Terminated |
| Nov 11, 2019 | Z-0778-2020, Z-0777-2020, Z-0776-2020Components 2mm shorter than indicated on the package label | Terminated |
| Oct 9, 2019 | Z-0534-2020Product labeled as STRM-NL-3550ST, NL LP Screw 3.5x50mm ST, lot M1151 may contain the incorrect screw type and length | Terminated |
Other HRS clearances
The newest 510(k)s for the same product code; their median review was 79 days; see every HRS device- Arthrex Plates, Screws and Pins, and WashersK263602 · Arthrex, Inc. · Oct 29, 2026
- Trochlear Salvage Plating SystemK261135 · Skeletal Dynamics, Inc. · Oct 26, 2026
- ANTHEM™ Fracture SystemK262779 · Globus Medical, Inc. · Oct 23, 2026
- Tempo Plating SystemK263407 · Pace Surgical · Oct 23, 2026
- Park's Pectus SystemK263269 · Tdm Co., Ltd. · Oct 15, 2026
- TDM Large Bone Plate and Screw SystemK261611 · Tdm Co., Ltd. · Oct 2, 2026
- Alians ForearmK262649 · Newclip Technics · Sep 28, 2026
- ProRib Fixation SystemK261892 · Costa Surgical, Inc. · Sep 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 14, 2018
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3030
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files