Bronchoscope (Flexible Or Rigid). 510(k) K182188, from Intuitive Surgical (Sunnyvale, CA). Product code EOQ, class 2.

As of Nov 17, 2026, K182188 is an FDA 510(k) held by Intuitive Surgical, substantially equivalent on Feb 14, 2019 after 185 days, with 5 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Feb 14, 2019
  • Review time: 185 days
  • Recalls: 5
  • Cited as a predicate by: 6 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Ion™ Endoluminal System (Model IF1000) assists the user in navigating a catheter and endoscopic tools in the pulmonary tract using endoscopic visualization of the tracheobronchial tree for diagnostic and therapeutic procedures. The Ion™ Endoluminal System enables fiducial marker placement. It does not make a diagnosis and is not for pediatric use.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K182188, newest first; each opens FDA's recall record
Jul 12, 2024Z-2720-2024There is a potential that the robotic system's instrument cart arm may contain screws that may potentially fail which could result in uncontrolled motion of a catheter inside a patient's airways.Open, Classified
Aug 6, 2021Z-0106-2022Potential for the sheath tip on the biopsy needle to separate from the sheath shaft.Terminated
Oct 23, 2020Z-0489-2021The distal tip ring of the fully articulating catheter may become dislodged during the procedure.Terminated
Jun 10, 2020Z-2949-2020One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or…Open, Classified
Jun 10, 2020Z-2950-2020One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or…Open, Classified

Other EOQ Clearances

The newest 510(k)s for the same product code; their median review was 229 days; see every EOQ device

Record

Type
Traditional
Received
Aug 13, 2018
Review panel
Ear, Nose, Throat
Regulation
21 CFR 874.4680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files