1688 Camera Control Unit and 4 more
Laparoscope, General & Plastic Surgery. 510(k) K182160, from Stryker (San Jose, CA). Product code GCJ, class 2.
As of Nov 18, 2026, K182160 is an FDA 510(k) held by Stryker, substantially equivalent on Nov 5, 2018 after 88 days, with 1 recall tied to it in FDA's recall database.
- Review time: 88 days
- Recalls: 1
- Cited as a predicate by: 5 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDF1688 4K Camera System with Advance Imaging Modality: The 1688 Video Camera is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery whenever a laparoscope/ endoscope/ arthroscope is indicated for use.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Stryker AIM Light Source and SafeLight cableK173866 · Stryker · Jan 18, 2018
- STRYKER INFRARED ILLUMINATION SYSTEM (IRIS) [AIM Light Source and IRIS Ureteral Kit]K151243 · Stryker Endoscopy · Aug 20, 2015
- Stryker Infrared Fluorescence (IRF) Imaging SystemK142310 · Stryker Endoscopy · Dec 2, 2014
- Stryker 1488 HD Video Camera With Infrared CompatibilityK132785 · Stryker Endoscopy · Dec 16, 2013
- PINPOINT Endoscopic Fluorescence Imaging SystemK150956 · Novadaq Technologies, Inc. · May 23, 2016
- PINPOINT Endoscopic Fluorescence Imaging SystemK161792 · Novadaq Technologies, Inc. · Sep 20, 2016
Cited as a predicate
5 later 510(k) summaries name this device as a predicate- Camera System (Camera Control Unit, Camera Head to be coupled to a fiberoptic scope), NIR FI Light SourceK221591 · Schoelly Fiberoptic GmbH · Feb 23, 2023
- Adavanced Imaging Modality System (1688 4K Camera System With Advanced Imaging Modality, L11 LED Light Source with…K202592 · Stryker · Oct 8, 2020
- 1688 Pendulum 4K Camera Head with Integrated Coupler, 1688 4K Camera Control Unit with Advanced Imaging ModalityK200310 · Stryker · Mar 6, 2020
- L10 LED Light Source with AIM, L11 LED Light Source with AIM, AIM SafeLight CableK192292 · Stryker · Nov 19, 2019
- L10 LED Light Source with AIM, L11 LED Light Source with AIMK191046 · Stryker · May 10, 2019
Recalls
Recalls FDA ties to K182160, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Started | Recall | Status |
|---|---|---|
| Jan 14, 2022 | Z-0728-2022A software defect in the camera control unit (CCU) will cause the image on the monitor to flip upside-down into an incorrect orientation. The potential of harms are conversion to open procedure, additional medical… | Open |
Other GCJ clearances
The newest 510(k)s for the same product code; their median review was 128 days; see every GCJ device- Swics™K263459 · Bloom Surgical, Inc. · Oct 29, 2026
- da Vinci 5 Handheld Camera Head (470080);da Vinci 5 Handheld Camera Laparoscope, 0 degree (470081);da Vinci 5 Handheld…K262455 · Intuitive Surgical, Inc. · Oct 14, 2026
- CuratOR EX2742K262397 · Eizo Corporation · Sep 11, 2026
- UNIDRIVE Select SystemK260884 · Karl Storz SE & CO. KG · Sep 1, 2026
- OEV-271MUH LCD Monitor;OEV-321MUH LCD Monitor;OEV-431MUH LCD MonitorK260972 · Sony Corporation · Aug 26, 2026
- VereSee (VER150, VER200, VER151, VER201, VER200C, and VER-100-103)K261139 · Freyja Healthcare, LLC · Aug 13, 2026
- EZ-Close™ (EZ Close Port Site Closing Device, EZ Close MAX Port Site Closing Device, EZ Close II Port Site Closing…K253541 · Medical Impact · Jul 24, 2026
- LED LCD Monitor (OEV-271H)K261614 · Lg Electronics.Inc · Jul 17, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 9, 2018
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.1500
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files