Arca
Skip to content
Tools
Tools

Mazor X System (Mazor X Stealth Edition)

Orthopedic Stereotaxic Instrument. 510(k) K182077, from Mazor Robotics , Ltd. (Caesarea, IL). Product code OLO, class 2.

As of Nov 18, 2026, K182077 is an FDA 510(k) held by Mazor Robotics , Ltd., substantially equivalent on Nov 2, 2018 after 93 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Nov 2, 2018
  • Review time: 93 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

Indications for Use The Mazor X is indicated for precise positioning of surgical instruments or spinal implants during general spinal and brain surgery. It may be used in open or minimally invasive or percutaneous procedures. Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C- Arms into volumetric 3D image.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate
  • Mazor XK200935 · Mazor Robotics , Ltd. · May 8, 2020

Recalls

Recalls FDA ties to K182077, newest first; each opens FDA's recall record
Dec 5, 2019Z-2330-2020There is a potential for the surgical system to detach from the operational room table unexpectedly. This could result in the surgical system falling onto the patient resulting in blunt injury [hemorrhage, hematoma,…Open, Classified
Jul 2, 2019Z-0721-2020The firm has become aware that the Snapshot Trackers were not manufactured in accordance with specifications and could potentially lead to navigational inaccuracies, which may result in patient injury, e.g. nerve damage…Terminated

Other OLO Clearances

The newest 510(k)s for the same product code; their median review was 91 days; see every OLO device

Record

Type
Traditional
Received
Aug 1, 2018
Review panel
Neurology
Regulation
21 CFR 882.4560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.