Enzymatic Method, Creatinine. 510(k) K182063, from Ortho-Clinical Diagnostics, Inc. (Rochester, NY). Product code JFY, class 2.

As of Nov 21, 2026, K182063 is an FDA 510(k) held by Ortho-Clinical Diagnostics, Inc., substantially equivalent on Oct 30, 2018 after 90 days, with 20 recalls tied to it in FDA's recall database.

  • Review time: 90 days
  • Recalls: 20
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

1. VITROS Chemistry Products CRBM Slides: Rx Only. For in vitro diagnostic use only. VITROS Chemistry Products CRBM Slides quantitatively measure carbamazepine (CRBM) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K182063, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 9, 2025Z-2039-2025a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date…Open
Dec 19, 2024Z-1006-2025The VITROS XT 7600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired…Open
Jan 26, 2024Z-1232-2024, Z-1233-2024A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules,…Open
Jan 26, 2024Z-1227-2024, Z-1228-2024A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.Open
Jan 22, 2024Z-1222-2024Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VITROS 3600 Immunodiagnostic Systems and VITROS 5600/XT 7600 Integrated Systems which may lead to erroneous test results.Open
Oct 10, 2023Z-0547-2024, Z-0546-2024During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen…Open
Aug 9, 2023Z-0008-2024Ortho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactured between 07-Apr-2022 (Date code: 040722) and 19-Dec-2022 (Date code: 121922) may not perform as intended due to an…Open
Mar 10, 2023Z-1555-2023MicroTip Pack Opener Assembly Potentially Not Removing or Replacing MicroTip Pack Caps on VITROS 5600 and XT 7600 Integrated Systems, results may be delayed results if a pack becomes unusableOpen
Jul 19, 2022Z-1675-2022, Z-1676-2022Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer.…Open
Mar 26, 2021Z-1489-2021Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpected color of the…Terminated
Jul 20, 2020Z-2878-2020, Z-2879-2020When processing MicroSlides, slides that utilize Slot 18 in the CM rotor may be mis-positioned. As a result, certain VITROS MicroSlides may produce erroneous results.Terminated
Jul 13, 2020Z-2873-2020Etrombopag Interference for VITROS Chemistry Products: TBIL Slides, VITROS BuBc Slides and VITROS XT TBIL-ALKP SlidesTerminated
Mar 20, 2019Z-1364-2019Potential for sample fluid to be dispensed to an incorrect position on the MicroSlide, potentially leading to erroneous assay results being reportedTerminated
Mar 11, 2019Z-1222-2019One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the…Terminated
Feb 20, 2019Z-1291-2019Luminometer Malfunction May Cause Inability to Process MicroWell Assays on VITROS SystemsTerminated

Other JFY clearances

The newest 510(k)s for the same product code; their median review was 185 days; see every JFY device

Record

Type
Traditional
Received
Aug 1, 2018
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1225

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files