Ingenia Ambition S and Ingenia Ambition X
System, Nuclear Magnetic Resonance Imaging. 510(k) K181479, from Philips Medical Systems Nederland B.V. (Best Noord-Brabant, NL). Product code LNH, class 2.
As of Nov 20, 2026, K181479 is an FDA 510(k) held by Philips Medical Systems Nederland B.V., substantially equivalent on Aug 3, 2018 after 59 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 59 days
- Recalls: 8
This system is a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities. Magnetic Resonance images represent the spatial distribution of protons or other nuclei with spin.
- Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4K173079 · Philips Medical Systems Nederland B.V. · Apr 4, 2018
| Jan 6, 2022 | Z-0975-2024Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field. | Open |
| Aug 11, 2021 | Z-2422-2021, Z-2416-2021When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment | Completed |
| Apr 23, 2021 | Z-1795-2021, Z-1796-2021The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving… | Open |
| Mar 25, 2019 | Z-1240-2019Magnet Energization Device (MED) can have a residual voltage on output terminal, may cause electrical shocks to personnel servicing the MED, MDD or connected cabling. | Terminated |
| Dec 21, 2018 | Z-0920-2019, Z-0921-2019There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. | Terminated |
- MAGNETOM Free.XLK262011 · Siemens Healthineers AG · Nov 5, 2026
- ECHELON SynergyK262902 · Fujifilm Corporation · Oct 9, 2026
- Vista AI ScanK260092 · Vista AI, Inc. · Oct 7, 2026
- SwiftSight-BodyCompositionK261612 · Airs Medical, Inc. · Oct 1, 2026
- APERTO Lucent MRI SystemK261283 · Fujifilm Corporation · Sep 16, 2026
- Embrace Vision Neonatal MRI SystemK261758 · Aspect Imaging, Ltd. · Sep 16, 2026
- MAGNETOM Free.Max; MAGNETOM Free.StarK260758 · Siemens Shenzhen Magnetic Resonance , Ltd. · Sep 1, 2026
- Swoop® Portable MR Imaging® SystemK262622 · Hyperfine, Inc. · Aug 19, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Abbreviated
- Received
- Jun 5, 2018
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1000
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files