Varian High Energy Linear Accelerator, UNIQUE, Varian Multileaf Collimator
Accelerator, Linear, Medical. 510(k) K181404, from Varian Medical Systems, Inc. (Palo Alto, CA). Product code IYE, class 2.
As of Nov 20, 2026, K181404 is an FDA 510(k) held by Varian Medical Systems, Inc., substantially equivalent on Jun 28, 2018 after 30 days, with 1 recall tied to it in FDA's recall database.
- Review time: 30 days
- Recalls: 1
The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. The UNIQUE is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
- Varian High Energy Linear AcceleratorK162476 · Varian Medical Systems, Inc. · Dec 12, 2016
- UniqueK132705 · Varian Medical Systems, Inc. · Oct 31, 2013
- Varian Multileaf CollimatorK133240 · Varian Medical Systems, Inc. · Jan 31, 2014
| Started | Recall | Status |
|---|---|---|
| Aug 12, 2019 | Z-1899-2020After a recent upgrade to the collimator software version 8.5, the firm became aware that the multi-leaf collimator leaves did not move during an arc treatment. The issue is related to a transient carriage primary-… | Terminated |
- AerFrame™ Alignment Laser Box (LBV2)K261865 · Stabilix, LLC · Sep 16, 2026
- EZFluence (RADEZ V2)K261718 · Radformation, Inc. · Aug 31, 2026
- MRIA Radiation Therapy Information System (MRIA)K260733 · Manteia Technologies Co., Ltd. · Aug 26, 2026
- Halcyon, Ethos Radiotherapy System (5.0)K261856 · Varian Medical Systems, Inc. · Aug 3, 2026
- ThinkQA 3K252937 · Dosisoft SA · Jun 12, 2026
- Luna 3DK253004 · Lap GmbH Laser Applikationen · Jun 4, 2026
- AerFrame Patient Immobilization System (AFPIS)K260792 · Stabilix, LLC · May 8, 2026
- Veriqa RT Epid 3DK252258 · Ptw-Freiburg Physikalisch-Technische-Werkstaetten Dr. Pychla · Apr 15, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- May 29, 2018
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files