Arca
Skip to content
Tools
Tools

Hamilton-T1, Hamilton-C1

Ventilator, Continuous, Facility Use. 510(k) K181216, from Hamilton Medical AG (Bonaduz, CH). Product code CBK, class 2.

As of Nov 19, 2026, K181216 is an FDA 510(k) held by Hamilton Medical AG, substantially equivalent on Aug 2, 2019 after 452 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 452 days
  • Recalls: 5

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The HAMILTON-C1 ventilator is intended to provide positive pressure ventilatory support or continuous flow of respiratory gases to adults and pediatrics, and optionally infants and neonates. Intended areas of use: • In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room room • During transfer of ventilated patients within the…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K181216, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 26, 2023Z-2526-2023, Z-2527-2023, Z-2525-2023Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of…Open
Jun 15, 2023Z-2139-2023, Z-2140-2023Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.Open

Other CBK clearances

The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device

Record

Type
Traditional
Received
May 7, 2018
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files