Trilogy Evo Universal
Ventilator, Continuous, Facility Use. 510(k) K181170, from Respironics, Inc. (Murrysville, PA). Product code CBK, class 2.
As of Nov 18, 2026, K181170 is an FDA 510(k) held by Respironics, Inc., substantially equivalent on Jul 18, 2019 after 442 days, with 13 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jul 18, 2019
- Review time: 442 days
- Recalls: 13
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Trilogy Evo Universal ventilator provides continuous or intermittent positive pressure ventilation for the care of individuals who require mechanical ventilation. Trilogy Evo Universal is intended for pediatric through adult patients weighing at least 2.5 kg.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Trilogy EvoK181166 · Respironics, Inc. · Jul 18, 2019
- Trilogy Series Ventilator With OximetryK111610 · Respironics, Inc. · Oct 14, 2011
- Vivo 60K160481 · Breas Medical AB · Nov 9, 2016
- Puritan Bennett 840 Series Ventilator SystemK151252 · Covidien LP · Mar 29, 2016
Recalls
Recalls FDA ties to K181170, newest first; each opens FDA's recall record| Mar 2, 2026 | Z-1649-2026In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths. | Open, Classified |
| Mar 2, 2026 | Z-1617-2026Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient. | Open, Classified |
| Oct 7, 2024 | Z-0292-2025Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements. | Open, Classified |
| Jul 16, 2024 | Z-2943-2024Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls. | Open, Classified |
| Jul 16, 2024 | Z-2944-2024Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls. | Open, Classified |
| Jul 16, 2024 | Z-2942-2024Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls. | Open, Classified |
| Mar 19, 2024 | Z-1505-2024Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power. | Open, Classified |
| Mar 19, 2024 | Z-1504-2024Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power. | Open, Classified |
| May 1, 2023 | Z-1652-2023The Trilogy Evo Universal ventilator provides continuous or intermittent positive pressure ventilation for the care of individuals who require mechanical ventilation. Trilogy Evo Universal is intended for pediatric… | Open, Classified |
| Mar 1, 2023 | Z-1289-2023The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value… | Open, Classified |
| Dec 22, 2021 | Z-0493-2022Philips has identified a specific lot of non-conforming material manufactured by one of its suppliers. A Philips supplier incorrectly used polyester-based polyurethane (PE-PUR) sound abatement foam, a non-conforming… | Open, Classified |
| Dec 22, 2021 | Z-0494-2022Philips has identified a specific lot of non-conforming material manufactured by one of its suppliers. A Philips supplier incorrectly used polyester-based polyurethane (PE-PUR) sound abatement foam, a non-conforming… | Open, Classified |
| Jul 26, 2021 | Z-2290-2021Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure… | Open, Classified |
Other CBK Clearances
The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device- Evita (V800); Evita (V600)K253647 · Draegerwerk AG & CO Kgaa · May 7, 2026
- Inbentus VersatileK252663 · Iconic Solutions by Murcia, SL · Apr 14, 2026
- VPAP Pediatric Face MaskK251888 · Hsiner Co., Ltd. · Mar 12, 2026
- Vivo 45 LSK232365 · Oconnell Regulatory Consultants, Inc. · Dec 19, 2025
- F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A); F&P OptiNIV…K252173 · Fisher & Paykel Healthcare, Ltd. · Sep 10, 2025
- SV600, SV800 VentilatorK243767 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Aug 28, 2025
- AVEA disposable expiratory filter/water trap (11790)K251133 · Telesair, Inc. · May 8, 2025
- AF531 Oro-Nasal SE Face MaskK243394 · Respironics, Inc. · Dec 30, 2024
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 2, 2018
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files