GammaPod
System, Radiation Therapy, Radionuclide. 510(k) K180571, from Xcision Medical Systems, LLC (Columbia, MD). Product code IWB, class 2.
As of Nov 20, 2026, K180571 is an FDA 510(k) held by Xcision Medical Systems, LLC, substantially equivalent on Apr 4, 2018 after 30 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 30 days
- Recalls: 2
- GammaPod - Model AK172706 · Xcision Medical Systems, LLC · Dec 22, 2017
| Nov 30, 2023 | Z-0750-2024Wrong structures imported into the GammaPod Treatment Planning System.-system allowed an RTSS generated from a previous CT scan to be selected | Open |
| Jul 3, 2019 | Z-2081-2019One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output. | Terminated |
- Leksell Gamma Knife® (Elekta Esprit); Leksell Gamma Knife® (Icon™); Leksell Gamma Knife® (Perfexion™)K250391 · Elekta Solutions AB · Jul 2, 2025
- Leksell GammaPlan (LGP)K232854 · Elekta Solutions AB · Feb 8, 2024
- Leksell Gamma Knife Perfexion, Leksell Gamma Knife Icon, Leksell Gamma Knife - Elekta Esprit (New model variant)K222047 · Elekta Solutions AB · Oct 26, 2022
- TaiChiCK203250 · Our United Corporation · Mar 12, 2021
- Akesis Galaxy RTiK203146 · Akesis, Inc. · Mar 5, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Mar 5, 2018
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files