Sonicision Reusable Generator, Sonicision Reusable Battery Pack, Sonicision Curved Jaw Dissectors (multiple lengths), Sonicision Battery Charger, Sonicision Battery Insertion Guide
Instrument, Ultrasonic Surgical. 510(k) K180149, from Covidien, LLC (Boulder, CO). Product code LFL, class U.
As of Nov 18, 2026, K180149 is an FDA 510(k) held by Covidien, LLC, substantially equivalent on Mar 26, 2018 after 66 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 66 days
- Recalls: 2
- Cited as a predicate by: 1 device
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Sonicision Cordless Ultrasonic Dissection Device Model SCD391, SCD396, SCT12, Scg, Scb, Scsk, Sonicision Sterilization TK101797 · Covidien, Formerly Valleylab, A Division of Tyco · Feb 24, 2011
- Sonicision Cordless Ultrasonic Dissection DeviceK153371 · Covidien, LLC · Mar 28, 2016
- Sonicision Cordless Ultrasonic DissectorK141371 · Covidien, Formerly Valleylab, A Division of Tyco H · Aug 7, 2014
Cited as a predicate
1 later 510(k) summary names this device as a predicate- Sonicision 7 Curved Jaw Cordless Ultrasonic Dissection DeviceK212301 · Covidien, LLC · Nov 15, 2021
Recalls
Recalls FDA ties to K180149, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Nov 12, 2021 | Z-0332-2022Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit | Open |
| Jun 23, 2021 | Z-2240-2021Potential for a manufacturing assembly error-may result in a non-functional audio indicator of device activation or a non-functional device, may cause unintended thermal burn, hemorrhage, or tissue damage and a delay of… | Terminated |
Other LFL clearances
The newest 510(k)s for the same product code; their median review was 158 days; see every LFL device- TORUS Ultrasonic Aspirator System; and accessories (TORUS Single-Use Tubing Sets, TORUS Single-Use Torque Wrenches,…K261361 · Carl Zeiss Meditec, AG · Oct 15, 2026
- Sonopet iQ Series III Console (5700-050-000); Sonopet iQ+ Straight Handpiece (5500-260-000); Sonopet iQ+ 10mm iQ+ Blade…K262244 · Stryker Instruments · Oct 8, 2026
- neXus Ultrasonic Surgical Aspirator SystemK261448 · Misonix, LLC · Jul 30, 2026
- CUSA® Clarity Ultrasonic Surgical Aspirator SystemK251162 · Integra LifeSciences Corporation · Oct 8, 2025
- SanAgile™ Ultrasonic Surgery Advanced Portable Controller (SA10); SanAgile™ Ultrasonic Surgery Advanced Dissector…K242894 · Shanghai Saints Sages Surgical Co., Ltd. · Jun 18, 2025
- Sonopet iQ Ultrasonic Aspirator System (5500-050-000)K243930 · Stryker Instruments · Mar 24, 2025
- Tenex 2nd Generation SystemK241700 · Trice Medical, Inc. · Nov 18, 2024
- CUSA® Clarity Ultrasonic Surgical Aspirator SystemK240493 · Integra LifeSciences Corporation · Jul 11, 2024
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 19, 2018
- Review panel
- Unknown
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files