Specular Microscope CEM-530
Microscope, Specular. 510(k) K173980, from Nidek Co., Ltd. (Gamagori, JP). Product code NQE, class 2.
As of Nov 20, 2026, K173980 is an FDA 510(k) held by Nidek Co., Ltd., substantially equivalent on Mar 14, 2018 after 75 days, with 1 recall tied to it in FDA's recall database.
- Review time: 75 days
- Recalls: 1
The NIDEK Specular Microscope CEM-530 is a non-contact ophthalmic microscope, optical pachymeter, and camera intended for examination of the corneal endothelium and for measurement of the thickness of the cornea.
- Specular Microscope CEM- 530K151706 · Nidek Co., Ltd. · Mar 17, 2016
- Konan Specular Microscope XivK120264 · Konan Medical, Inc. · Apr 11, 2012
| Started | Recall | Status |
|---|---|---|
| Feb 25, 2019 | Z-0771-2020The electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2) | Terminated |
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 29, 2017
- Review panel
- Ophthalmic
- Regulation
- 21 CFR 886.1850
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files