Arca
Skip to content

Triathlon Total Knee System

Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer. 510(k) K173849, from Stryker Orthopaedics (Mahwah, NJ). Product code MBH, class 2.

As of Nov 17, 2026, K173849 is an FDA 510(k) held by Stryker Orthopaedics, substantially equivalent on Feb 2, 2018 after 45 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Feb 2, 2018
  • Review time: 45 days
  • Recalls: 3
  • Cited as a predicate by: 4 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

General Total Knee Arthroplasty (TKR) Indications: • Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis. • Post-traumatic loss of knee joint configuration and function.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K173849, newest first; each opens FDA's recall record
Feb 6, 2024Z-1260-2024Potential packaging breaches of inner blister and outer sterile blister.Open, Classified
Feb 6, 2024Z-1259-2024Potential packaging breaches of inner blister and outer sterile blister.Open, Classified
Apr 13, 2023Z-1653-2023Mislabeled:Labeled Catalog Number: 5531-G-312-E, TRIATHLON X3 TIBIAL BEARING INSERT CS #3 12MM, may contain Catalog Number: 5532-G-413-E, TRIATHLON X3 TIBIAL BEARING INSERT PS #4 13MM, and vice versaOpen, Classified

Other MBH Clearances

The newest 510(k)s for the same product code; their median review was 153 days; see every MBH device

Record

Type
Traditional
Received
Dec 19, 2017
Review panel
Orthopedic
Regulation
21 CFR 888.3565

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.