Arca
Skip to content
Tools
Tools

Quantum Pump Console

Console, Heart-Lung Machine, Cardiopulmonary Bypass. 510(k) K173834, from Spectrum Medical , Ltd. (Gloucester, GB). Product code DTQ, class 2.

As of Nov 28, 2026, K173834 is an FDA 510(k) held by Spectrum Medical , Ltd., substantially equivalent on May 25, 2018 after 158 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 158 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Quantum Pump Console is indicated for use for up to 6 hours in cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K173834, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 13, 2023Z-0224-2025The ratchet and pawl had an out of tolerance geometry which exhibited as the left bobbin would not lock.Completed
May 29, 2019Z-2238-2019Potential for Quantum Pump Console, part of the Quantum Perfusion Systems, to unexpectedly shut down while in use.Terminated

Other DTQ clearances

The newest 510(k)s for the same product code; their median review was 219 days; see every DTQ device

Record

Type
Traditional
Received
Dec 18, 2017
Review panel
Cardiovascular
Regulation
21 CFR 870.4220

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc