Arca
Skip to content
Tools
Tools

Medtronic O-arm O2 Imaging System

Interventional Fluoroscopic X-Ray System. 510(k) K173664, from Medtronic, Inc. (Littleton, MA). Product code OWB, class 2.

As of Nov 18, 2026, K173664 is an FDA 510(k) held by Medtronic, Inc., substantially equivalent on Dec 29, 2017 after 30 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Dec 29, 2017
  • Review time: 30 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The O- Drm O2 Imaging System is a mobile x-ray system, designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighing 60 lbs or greater and having an abdominal thickness greater than 16cm, and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K173664, newest first; each opens FDA's recall record
Nov 15, 2019Z-0767-2020Due to a manufacturing issue affecting particular O-arm" O2 Imaging Systems, Medtronic has determined that the gantry tractor motor drive belt may loosen over an extended number of uses at a higher frequency in these…Terminated
Sep 17, 2018Z-0404-2019O-arm 02 Imaging Systems correction to software version 4.1.0; new version of the user manual and an Image Acquisition System ("IAS") product label; replacing blown fuse of the new AC power input circuitTerminated

Other OWB Clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Special
Received
Nov 29, 2017
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.