ACUSON AcuNav Volume Intracardiac Echocardiography Catheter
Catheter, Ultrasound, Intravascular. 510(k) K173618, from Siemens Medical Solutions USA, Inc. (Moutain View, CA). Product code OBJ, class 2.
As of Nov 20, 2026, K173618 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Dec 19, 2017 after 27 days, with 1 recall tied to it in FDA's recall database.
- Review time: 27 days
- Recalls: 1
- Cited as a predicate by: 3 devices
The catheter is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
- AcuNav Volume ICE CatheterK163478 · Siemens Medical Solutions USA, Inc. · Feb 9, 2017
- ACUSON AcuNav Volume Intracardiac Echocardiography(ICE) CatheterK203726 · Siemens Medical Solutions USA, Inc. · Jan 19, 2021
- VeriSight Intracardiac Echocardiography (ICE) Catheter, VeriSight Pro ICE CatheterK200812 · Philips Image Guided Therapy Corporation · Sep 2, 2020
- ACUSON SC2000 Diagnostic Ultrasound SystemK181098 · Siemens Medical Solutions USA, Inc. · May 31, 2018
| Started | Recall | Status |
|---|---|---|
| May 22, 2020 | Z-0090-2021Due to adverse events received by the firm for left atrial perforations and Atrioventricular Node block. | Terminated |
- ACUNAV Luxera 4D ICE CatheterK261696 · Siemens Medical Solutions USA, Inc. · Sep 10, 2026
- VERAFEYE Imaging Catheter (VF-VIC-002)K261849 · Luma Vision Limited · Jul 24, 2026
- Novasight Hybrid SystemK252945 · Conavi Medical, Inc. · Apr 17, 2026
- OptiCross Coronary Imaging Catheter (H749518110); OptiCross 6 Coronary Imaging Catheter (H7495181160); OptiCross HD…K260816 · Boston Scientific · Apr 8, 2026
- Visions® PV .014P RX Digital IVUS Catheter ; Visions® PV .018 Digital IVUS CatheterK253399 · Volcano Corporation (Dba Philips Image Guided Therapy Device · Oct 28, 2025
- XC11 ICE System, USAK250955 · Yorlabs, Inc. · Oct 22, 2025
- TINGSN FINDERS 2 Color Doppler Diagnostic Ultrasound SystemK250913 · Jiangsu Tingsn Technology Co., Ltd. · Jul 19, 2025
- ViewFlex™ Xtra ICE Catheter; ViewFlex™ Eco Reprocessed ICE Catheter; Advisor™ HD Grid Mapping Catheter, Sensor…K251211 · Abbott Medical · May 23, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Nov 22, 2017
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1200
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files