GYN-Pump PH304
Insufflator, Hysteroscopic. 510(k) K173489, from W.O.M. World of Medicine GmbH (Berlin, DE). Product code HIG, class 2.
As of Nov 20, 2026, K173489 is an FDA 510(k) held by W.O.M. World of Medicine GmbH, substantially equivalent on Mar 28, 2018 after 135 days, with 1 recall tied to it in FDA's recall database.
- Review time: 135 days
- Recalls: 1
The GYN-Pump PH304 is a combined suction and irrigation pump for use in hysteroscopic and laparoscopic interventions. During diagnostic and operative hysteroscopy, the GYN-Pump PH304 is intended to provide liquid distension of the uterus and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus.
- Aquilex Fluid Control System H112K112642 · W.O.M. World of Medicine AG · Feb 6, 2012
- LAP-Pump PP110K163320 · W.O.M. World of Medicine GmbH · Mar 21, 2017
- Hysteroscopy Pump HM6K123732 · W.O.M. World of Medicine AG · Apr 4, 2013
| Started | Recall | Status |
|---|---|---|
| Jun 18, 2021 | Z-2561-2021When the display of inflow volume to the uterus reaches 32450 ml, the fluid management system control unit display will freeze because the internal software calculation threshold is reached. The outflow measurement will… | Open |
- Endomat Select (UP220)K261706 · Karl Storz Se & Co. KG · Jul 16, 2026
- Asurys Fluid Management SystemK251273 · Boston Scientific Corporation · Mar 27, 2026
- Fluent Pro Fluid Management System (FLT-200); Fluent Pro Fluid Management System Disposable Procedure Kit (6-pack)…K240886 · Hologic, Inc. · Jul 25, 2024
- X-FLO Fluid Management SystemK210628 · Thermedx, LLC · Nov 4, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 13, 2017
- Review panel
- Obstetrics/Gynecology
- Regulation
- 21 CFR 884.1700
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files