Ingenia Elition S and Ingenia Elition X
System, Nuclear Magnetic Resonance Imaging. 510(k) K173451, from Philips Medical Systems Nederland B.V. (Best Noord-Brabant, NL). Product code LNH, class 2.
As of Nov 18, 2026, K173451 is an FDA 510(k) held by Philips Medical Systems Nederland B.V., substantially equivalent on Mar 20, 2018 after 134 days, with 14 recalls tied to it in FDA's recall database.
- Substantially equivalent: Mar 20, 2018
- Review time: 134 days
- Recalls: 14
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3K163116 · Philips Medical Systems Nederland B.V. · Jan 6, 2017
Recalls
Recalls FDA ties to K173451, newest first; each opens FDA's recall record| Jun 12, 2023 | Z-2396-2023The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system. | Open, Classified |
| Jun 12, 2023 | Z-2399-2023During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does… | Open, Classified |
| Jun 12, 2023 | Z-2395-2023The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system. | Open, Classified |
| Jun 12, 2023 | Z-2398-2023During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does… | Open, Classified |
| Nov 22, 2022 | Z-0460-2023Gradient Coil may act as a heat source with a potential to produce smoke and/or fire. | Open, Classified |
| Nov 22, 2022 | Z-0461-2023Gradient Coil may act as a heat source with a potential to produce smoke and/or fire. | Open, Classified |
| Nov 22, 2022 | Z-0463-2023Gradient Coil may act as a heat source with a potential to produce smoke and/or fire. | Open, Classified |
| Nov 22, 2022 | Z-0462-2023Gradient Coil may act as a heat source with a potential to produce smoke and/or fire. | Open, Classified |
| Nov 22, 2022 | Z-0459-2023Gradient Coil may act as a heat source with a potential to produce smoke and/or fire. | Open, Classified |
| Jan 6, 2022 | Z-0974-2024Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field. | Open, Classified |
| Aug 11, 2021 | Z-2419-2021When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment | Completed |
| Aug 11, 2021 | Z-2418-2021When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment | Completed |
| Jul 30, 2021 | Z-2347-2021Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury | Terminated |
| Jul 30, 2021 | Z-2344-2021Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury | Terminated |
Other LNH Clearances
The newest 510(k)s for the same product code; their median review was 152 days; see every LNH device- ECHELON SynergyK262902 · Fujifilm Corporation · Oct 9, 2026
- Vista AI ScanK260092 · Vista AI, Inc. · Oct 7, 2026
- SwiftSight-BodyCompositionK261612 · Airs Medical, Inc. · Oct 1, 2026
- APERTO Lucent MRI SystemK261283 · Fujifilm Corporation · Sep 16, 2026
- Embrace Vision Neonatal MRI SystemK261758 · Aspect Imaging, Ltd. · Sep 16, 2026
- MAGNETOM Free.Max; MAGNETOM Free.StarK260758 · Siemens Shenzhen Magnetic Resonance , Ltd. · Sep 1, 2026
- Swoop® Portable MR Imaging® SystemK262622 · Hyperfine, Inc. · Aug 19, 2026
- MagicScan-1.5, GreenMR Serenity 1.5TK254250 · Xingaoyi Medical Equipment Co.,Ltd · Aug 17, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Abbreviated
- Received
- Nov 6, 2017
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1000
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files