ProxiDiagnost N90

Interventional Fluoroscopic X-Ray System. 510(k) K173433, from Philips Medical Systems Dmc GmbH (Hamburg, DE). Product code OWB, class 2.

As of Nov 20, 2026, K173433 is an FDA 510(k) held by Philips Medical Systems Dmc GmbH, substantially equivalent on Feb 5, 2018 after 95 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 95 days
  • Recalls: 7
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

ProxiDiagnost N90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K173433, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 3, 2023Z-1587-2023Philips has identified a compliance issue where ProxiDiagnost N90 R.1.0, CombiDiagnost R90 R.1.0 and CombiDiagnost R90 R1.1 systems are missing a certification label required by the Food and Drug Administration (FDA).Open
Oct 21, 2022Z-1477-2023Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop…Open
Oct 21, 2021Z-1356-2023If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell…Open
Sep 22, 2021Z-0240-2022The Aleva monitor requires certain labels to be applied, some of which are required by FDA Electronic Product Radiation Control performance standards. This includes the Radiation Warning Label [21CFR 1020.30 (j)] and…Open
Dec 8, 2020Z-0727-2021After the user selects the Lock-in command, the kV and mA values are not locked immediately and can adjust if the collimator shutter position is changed directly after the Lock-in command.Terminated
Apr 23, 2020Z-2256-2020The thermo switches in the three-phase transformer, which normally aid in powering down the System Power Distribution Unit, may be incorrectly installed and not working. If the transformer overheats due to a first…Terminated
Nov 20, 2019Z-3023-2020Unexpected increase in peak tube potential (kV) will lead to an increased patient radiation dose.Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
Nov 2, 2017
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files