Arca
Skip to content
Tools
Tools

Mammomat Revelation

Full Field Digital, System, X-Ray, Mammographic. 510(k) K173408, from Siemens Medi Cal Solutions, Inc. (Malvern, PA). Product code MUE, class 2.

As of Nov 20, 2026, K173408 is an FDA 510(k) held by Siemens Medi Cal Solutions, Inc., substantially equivalent on Mar 21, 2018 after 140 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 140 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The MAMMOMAT Revelation is intended to be used for mammography exams, screening, diagnostics, biopsies and dual energy procedures under the supervision of medical professionals. The Mammography images can be interpreted by either hard copy film or soft copy workstation.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K173408, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 7, 2020Z-2388-2020If an examination is interrupted when the InSpect Projection View (PV) is active and the vacuum biopsy system is being used, the tube arm's manual movement range is decreased from +/-15 deg. to +/- 6 deg. This…Terminated
Jul 23, 2019Z-0160-2020An error can occur when performing a biopsy using the InSpect function. The acquisition workstation may become unresponsive to normal user interaction, can only take place in rare casesTerminated

Other MUE clearances

The newest 510(k)s for the same product code; their median review was 209 days; see every MUE device

Record

Type
Traditional
Received
Nov 1, 2017
Review panel
Radiology
Regulation
21 CFR 892.1715

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files