Arca
Skip to content

Ingenia Elition S and Ingenia Elition X R5.4

System, Nuclear Magnetic Resonance Imaging. 510(k) K173356, from Philips Medical Systems Nederland B.V. (Best Noord-Brabant, NL). Product code LNH, class 2.

As of Nov 18, 2026, K173356 is an FDA 510(k) held by Philips Medical Systems Nederland B.V., substantially equivalent on Mar 6, 2018 after 132 days, with 13 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Mar 6, 2018
  • Review time: 132 days
  • Recalls: 13

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

This system is a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities. Magnetic Resonance images represent the spatial distribution of protons or other nuclei with spin.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K173356, newest first; each opens FDA's recall record
Jun 12, 2023Z-2396-2023The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.Open, Classified
Jun 12, 2023Z-2399-2023During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does…Open, Classified
Jun 12, 2023Z-2395-2023The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.Open, Classified
Jun 12, 2023Z-2398-2023During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does…Open, Classified
Nov 22, 2022Z-0460-2023Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.Open, Classified
Nov 22, 2022Z-0461-2023Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.Open, Classified
Nov 22, 2022Z-0463-2023Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.Open, Classified
Nov 22, 2022Z-0462-2023Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.Open, Classified
Nov 22, 2022Z-0459-2023Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.Open, Classified
Jan 6, 2022Z-0974-2024Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.Open, Classified
Dec 21, 2018Z-0922-2019There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.Terminated
Dec 21, 2018Z-0923-2019There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.Terminated
Dec 3, 2018Z-0694-2019During automatic tabletop movement, pressing and holding the Stop Table button may lead to an unexpected restart of the table movement.Terminated

Other LNH Clearances

The newest 510(k)s for the same product code; their median review was 152 days; see every LNH device

Record

Type
Abbreviated
Received
Oct 25, 2017
Review panel
Radiology
Regulation
21 CFR 892.1000

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.