Astral 100/150
Ventilator, Continuous, Facility Use. 510(k) K172875, from Resmed, Ltd. (Bella Vista, AU). Product code CBK, class 2.
As of Nov 20, 2026, K172875 is an FDA 510(k) held by Resmed, Ltd., substantially equivalent on Apr 26, 2018 after 217 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 217 days
- Recalls: 2
The Astral 100/150 provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation. The iVAPS mode with optional AutoEPAP is intended for patients weighing more than 66lb (30kg). The Astral device is intended to be used in home, institution/hospital and portable applications for both invasive and non- invasive ventilation.
- Astral 100/150K152068 · Resmed, Ltd. · May 13, 2016
| Jun 25, 2026 | Z-2735-2026Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy. | Open |
| Sep 13, 2023 | Z-0111-2024If ventilator is on internal battery, not intended to serve as a primary power source, low/critically low battery alarms will sound, but a fault leads to sudden power loss. If power fails, then ventilation stops and a… | Open |
- Evita (V800); Evita (V600)K253647 · Draegerwerk AG & CO Kgaa · May 7, 2026
- Inbentus VersatileK252663 · Iconic Solutions by Murcia, SL · Apr 14, 2026
- VPAP Pediatric Face MaskK251888 · Hsiner Co., Ltd. · Mar 12, 2026
- Vivo 45 LSK232365 · Oconnell Regulatory Consultants, Inc. · Dec 19, 2025
- F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A); F&P OptiNIV…K252173 · Fisher & Paykel Healthcare, Ltd. · Sep 10, 2025
- SV600, SV800 VentilatorK243767 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Aug 28, 2025
- AVEA disposable expiratory filter/water trap (11790)K251133 · Telesair, Inc. · May 8, 2025
- AF531 Oro-Nasal SE Face MaskK243394 · Respironics, Inc. · Dec 30, 2024
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 21, 2017
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files