MEVION S250i Proton Beam Radiation Therapy Device

System, Radiation Therapy, Charged-Particle, Medical. 510(k) K172848, from Mevion Medical Systems (Littleton, MA). Product code LHN, class 2.

As of Nov 19, 2026, K172848 is an FDA 510(k) held by Mevion Medical Systems, substantially equivalent on Dec 27, 2017 after 99 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 99 days
  • Recalls: 5
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The MEVION S250i is intended to deliver proton radiation therapy treatment to patients with localized tumors or other conditions susceptible to treatment by radiation.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K172848, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 1, 2022Z-0827-2022Malfunction of the hand pendant controlling movement of the therapy couch and nozzle may result in unintentional motion. Patient injury is possible if they come in contact with the nozzle or other objects in the…Open
Oct 19, 2020Z-0485-2021Treatment beam information disappears on Treatment Console screen while beam delivery continues potential for harm to a patient could occurTerminated
May 15, 2020Z-2406-2020Couch Correction moves sent after using a 3D CT scan are partially lost if the previous correction sent or recorded had some of the Couch Correction checkboxes turned offTerminated
Oct 1, 2018Z-0705-2019Treatment is allowed to continue (via partials) in situations where dDose1 interlock has trippedTerminated
Jun 29, 2018Z-2746-2018QFix has advised Mevion Medical Systems that they are identifying unauthorized combinations of QFix Couch Tops with non-QFix devices in clinical use. The kVue Oen Touch latch consists of a precision latch mechanism that…Terminated

Other LHN clearances

The newest 510(k)s for the same product code; their median review was 190 days; see every LHN device

Record

Type
Traditional
Received
Sep 19, 2017
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files