Immunoassay Method, Troponin Subunit. 510(k) K172783, from Beckman Coulter, Inc. (Chaska, MN). Product code MMI, class 2.
As of Nov 20, 2026, K172783 is an FDA 510(k) held by Beckman Coulter, Inc., substantially equivalent on Jun 12, 2018 after 270 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 270 days
- Recalls: 2
Access hsTnI is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma using the UniCel DxI Access Immunoassay Systems to aid in the diagnosis of myocardial infarction (MI).
- Access Accutni+3 Reagent and Access Accutn1+3 Calibrators for Use On The Access 2 Immunoassay SystemK121214 · Beckman Coulter, Inc. · Jun 14, 2013
| Nov 18, 2024 | Z-0727-2025Beckman Coulter, Inc. is recalling their Access hsTnI assay, a medical device by correction. The reason for the recall is: high troponin antigen (>55,000 pg/mL) can contaminate an analyzers probe and cause inter-assay… | Open |
| Aug 9, 2021 | Z-0032-2022Through customer feedback, testing a sample with cardiac troponin I concentration that are >270,000 pg/mL may cause carryover into the hsTnl reagent pack. The carryover is caused by sample-to-pack carryover… | Open |
- i-STAT hs-TnI cartridge with the i-STAT Alinity SystemK253051 · Abbott Point of Care, Inc. · Jun 11, 2026
- VITROS Immunodiagnostic Products hs Troponin I Reagent PackK252393 · Ortho-Clinical Diagnostics, Inc. · Oct 29, 2025
- Access hsTnIK243483 · Beckman Coulter, Inc. · Aug 1, 2025
- Access hsTnIK242870 · Beckman Coulter, Inc. · Jun 16, 2025
- i-STAT hs-TnI cartridge with the i-STAT 1 SystemK240984 · Abbott Point of Care, Inc. · Jan 3, 2025
- Atellica® IM High-Sensitivity Troponin I (TnIH)K241165 · Siemens Healthcare Diagnostics, Inc. · Jul 25, 2024
- PATHFAST®hs-cTnI-IIK231974 · Phc Corporation · Mar 20, 2024
- Access hsTnIK222881 · Beckman Coulter, Inc. · Dec 18, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 15, 2017
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1215
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files