Olympus URF-P6/P6R
Ureteroscope And Accessories, Flexible/Rigid. 510(k) K172298, from Olympus Medical Systems Corp. (Hachioji-Shi, JP). Product code FGB, class 2.
As of Nov 18, 2026, K172298 is an FDA 510(k) held by Olympus Medical Systems Corp., substantially equivalent on Mar 29, 2018 after 241 days, with 5 recalls tied to it in FDA's recall database.
- Substantially equivalent: Mar 29, 2018
- Review time: 241 days
- Recalls: 5
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThis instrument has been designed to be used with an Olympus light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct).
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- URF-P2 Ureterorenofiberscope/CholedochofiberscopeK912120 · Olympus Corp. · Aug 7, 1991
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- Uretero-Reno Videoscope Olympus URF-V3/V3R, Uretero-Reno Fiberscope Olympus URF-P7/P7RK181451 · Olympus Medical Systems Corp. · Jan 16, 2019
Recalls
Recalls FDA ties to K172298, newest first; each opens FDA's recall record| Sep 25, 2023 | Z-0347-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for… | Open, Classified |
| Apr 20, 2022 | Z-1263-2022Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with… | Open, Classified |
| Apr 20, 2022 | Z-1264-2022Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with… | Open, Classified |
| Jan 17, 2018 | Z-2520-2018Breakage of the endoscope's insertion tube bending section during surgical procedures. | Terminated |
| Jan 17, 2018 | Z-2519-2018Breakage of the endoscope's insertion tube bending section during surgical procedures. | Terminated |
Other FGB Clearances
The newest 510(k)s for the same product code; their median review was 140 days; see every FGB device- Single-use Digital Flexible Ureteroscope (GY-UR6.3); Single-use Digital Flexible Ureteroscope (GY-UR6.3A)K262437 · Zhejiang Geyi Medical Instrument Co., Ltd. · Oct 29, 2026
- Vein360 Reprocessed LithoVue Digital Flexible Ureteroscope (Standard Deflection); Vein360 Reprocessed LithoVue Digital…K260708 · Vein 360 · Oct 20, 2026
- CVAC® Aspiration SystemK261447 · Calyxo, Inc. · Sep 15, 2026
- Single-Use Video Flexible Ureterorenoscope (RP-U-C06R9, RP-U-C06R3, RP-U-C06S9, RP-U-C06S3)K262488 · Guangzhou Red Pine Medical Instrument Co., Ltd. · Sep 9, 2026
- LithoVue Elite Digital Flexible Ureteroscope SystemK261153 · Boston Scientific Corporation · Aug 26, 2026
- Single-use Flexible Ureteroscope (Single-use Flexible Ureteroscope: SUV-5C-B)K262552 · Scivita Medical Technology Co., Ltd. · Aug 14, 2026
- Zamenix™ PK253505 · Roen Surgical, Inc. · Aug 7, 2026
- BubbleView Single-Use Digital Flexible Cystoscope (C38-CS, C38-CRS, C50-C, C50-CR, C50-CB, C50-CBR)K260516 · MacroLux Medical Technology Co., Ltd. · Jul 20, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 31, 2017
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.1500
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files