Ureteroscope And Accessories, Flexible/Rigid. 510(k) K172298, from Olympus Medical Systems Corp. (Hachioji-Shi, JP). Product code FGB, class 2.

As of Nov 18, 2026, K172298 is an FDA 510(k) held by Olympus Medical Systems Corp., substantially equivalent on Mar 29, 2018 after 241 days, with 5 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Mar 29, 2018
  • Review time: 241 days
  • Recalls: 5
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

This instrument has been designed to be used with an Olympus light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct).

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K172298, newest first; each opens FDA's recall record
Sep 25, 2023Z-0347-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for…Open, Classified
Apr 20, 2022Z-1263-2022Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with…Open, Classified
Apr 20, 2022Z-1264-2022Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with…Open, Classified
Jan 17, 2018Z-2520-2018Breakage of the endoscope's insertion tube bending section during surgical procedures.Terminated
Jan 17, 2018Z-2519-2018Breakage of the endoscope's insertion tube bending section during surgical procedures.Terminated

Other FGB Clearances

The newest 510(k)s for the same product code; their median review was 140 days; see every FGB device

Record

Type
Traditional
Received
Jul 31, 2017
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1500

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.