Aquilex Fluid Control System AQL-100S
Insufflator, Hysteroscopic. 510(k) K172040, from W.O.M. World of Medicine GmbH (Berlin, DE). Product code HIG, class 2.
As of Nov 19, 2026, K172040 is an FDA 510(k) held by W.O.M. World of Medicine GmbH, substantially equivalent on Feb 5, 2018 after 215 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 215 days
- Recalls: 2
The Aquilex® Fluid Control System AQL-100S is intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopy and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus.
- Aquilex Fluid Control System H112K112642 · W.O.M. World of Medicine AG · Feb 6, 2012
- Hysteroscopy Pump HM6K123732 · W.O.M. World of Medicine AG · Apr 4, 2013
| Jul 17, 2023 | Z-2629-2023The display of inflow volume can reach its limit of 30,000 ml during long procdeure and the inflow volume display will freeze at the maximum value while the deficit will start counting backwards until 0 ml is reached… | Open |
| Aug 14, 2019 | Z-2767-2019The deficit displayed by the pump can differ from the real deficit, leading potentially to patient risk. | Terminated |
- Endomat Select (UP220)K261706 · Karl Storz Se & Co. KG · Jul 16, 2026
- Asurys Fluid Management SystemK251273 · Boston Scientific Corporation · Mar 27, 2026
- Fluent Pro Fluid Management System (FLT-200); Fluent Pro Fluid Management System Disposable Procedure Kit (6-pack)…K240886 · Hologic, Inc. · Jul 25, 2024
- X-FLO Fluid Management SystemK210628 · Thermedx, LLC · Nov 4, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Jul 5, 2017
- Review panel
- Obstetrics/Gynecology
- Regulation
- 21 CFR 884.1700
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files