Arca
Skip to content
Tools
Tools

Aquilex Fluid Control System AQL-100S

Insufflator, Hysteroscopic. 510(k) K172040, from W.O.M. World of Medicine GmbH (Berlin, DE). Product code HIG, class 2.

As of Nov 19, 2026, K172040 is an FDA 510(k) held by W.O.M. World of Medicine GmbH, substantially equivalent on Feb 5, 2018 after 215 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 215 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Aquilex® Fluid Control System AQL-100S is intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopy and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K172040, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 17, 2023Z-2629-2023The display of inflow volume can reach its limit of 30,000 ml during long procdeure and the inflow volume display will freeze at the maximum value while the deficit will start counting backwards until 0 ml is reached…Open
Aug 14, 2019Z-2767-2019The deficit displayed by the pump can differ from the real deficit, leading potentially to patient risk.Terminated

Other HIG clearances

The newest 510(k)s for the same product code; their median review was 247 days; see every HIG device

Record

Type
Special
Received
Jul 5, 2017
Review panel
Obstetrics/Gynecology
Regulation
21 CFR 884.1700

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.