MAGEC® System
Growing Rod System- Magnetic Actuation. 510(k) K171791, from Nuvasive Specialized Orthopedics, Incorporated (Alisa Viejo, CA). Product code PGN, class 2.
As of Nov 21, 2026, K171791 is an FDA 510(k) held by Nuvasive Specialized Orthopedics, Incorporated, substantially equivalent on Aug 31, 2017 after 76 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 76 days
- Recalls: 2
- Cited as a predicate by: 1 device
The MAGEC® System is intended for skeletally immature patients less than 10 years of age with severe progressive spinal deformities (e.g., Cobb angle of 30 degrees or more; thoracic spine height less than 22 cm) associated with or at risk of Thoracic Insufficiency Syndrome (TIS). TIS is defined as the inability of the thorax to support normal respiration or lung growth.
- MAGEC 2 Spinal Bracing and Distraction SystemK150885 · Ellipse Technologies, Incorporated · May 29, 2015
- MAGEC® Spinal Bracing and Distraction SystemK161751 · Nuvasive Specialized Orthopedics, Inc. · Sep 2, 2016
- MAGEC® Spinal Bracing and Distraction SystemK160352 · Nuvasive Specialized Orthopedics, Inc. · Oct 21, 2016
- Ellipse Magec Spinal Bracing and Distraction SystemK140178 · Ellipse Technologies, Inc. · Feb 27, 2014
- MAGEC® SystemK201543 · Nuvasive Specialized Orthopedics, Inc. · Jul 30, 2020
| Apr 3, 2025 | Z-1657-2025Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm. | Open |
| Feb 13, 2020 | Z-1898-2020Post-implantation separation of an actuator end cap component that may expose internal components of the actuator, which could lead to hastened degeneration of the internal components and egress of Titanium alloy wear… | Terminated |
- MAGEC Spinal Bracing and Distraction SystemK233593 · Nuvasive Specialized Orthopedics, Incorporated · Feb 6, 2024
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 16, 2017
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3070
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files